HPLC Columns - Zorbax Columns
Back
Agilent ZORBAX StableBond Column Specifications and Options
 
ZORBAX StableBond SB-Aq
Agilent ZORBAX SB-Aq is made by chemically bonding a reversed-phase proprietary alkyl group to special ZORBAX silica.
 
Particle Size (µm) Length (mm) ID Specifications Method Development Notes Working with LC/MS
1.8 µm particles (both RRHD, stable to 1200 bar, and RRHT, stable to 600 bar)
2.7 µm Poroshell 120 (superficially porous)
3.5 µm
5 µm
7 µm
20 - 250 1.0
2.1
3.0
4.6
Capillary and Prep
Custom options available
Non-endcapped
Pore Size: 80Å
Surface area: 180 m2/g
pH: 1.0 - 8.0
Max temperature: 80 °C
Carbon load: proprietary
Start with 5% methanol or acetonitrile in water as the initial solvent, and 100% methanol or acetonitrile as the final solvent. We recommend adding 0.1% formic acid in both A and B bottles. This column can be used with 100% aqueous. If using LC/MS, we recommend starting with 5 - 10 mM ammonium formate, ammonium acetate, ammonium hydroxide, 0.1% acetic acid or 0.1% formic acid. We recommend against using ammonium bicarbonate.
 
For more information click here
Back
 
 
 
 
Copyright © LCGC. All Right Reserved.